Surrounding |
Area |
Clean & free from dust |
Temp |
20 ℃ ~ 35 ℃ (Recommended using AC) |
Level |
Horizontally leveled |
Clearance Space |
Right & left Side |
Minimum 500mm |
Back Side |
Minimum 500mm |
Top Side |
Minimum 500mm |
Door Open Area |
Greater than doors width size of equipment |
Refrigeration unit (for Walk in type chamber) |
Distance between equipment and compressor unit should not exceed 45 feet |
3“ diameter hole required at wall for passing copper pipe of split type refrigeration unit. |
Proper shade is required to protect compressor unit from rain water. |
Required stabilized power supply for equipment |
For non walk-in type |
Single phase, 230+10 VAC, 50 Hz, 20 Amp with proper earthing |
For walk-in type |
Three phase, 440+10 VAC, 50 Hz, 32 Amp with proper earthing |
Required UPS for PLC & Control system |
Single phase, 230+10 VAC, 50 Hz, 250 VA |
Required LAN port [I/O] for communication of equipment and software |
Information outlet [I/O] for each equipment and distance between equipment & port should not exceed 2.5 meters. |
Drainage |
Level |
Concealed (below ground level) |
Continuous water supply
(For Temperature & Humidity base Equipment) |
Pipe specification |
3“ diameter PVC |
Water quality |
Dust free Demineralized or soft water |
Pressure |
1 kg/cm ~ 1 Bar ~ 15 psi approx |
Valve diameter |
½ “BSP |